Cook Medical Inc. has stopped shipments of its Zilver PTX drug-eluting stents and is recalling more than 40,000 units worldwide due to reports that the device delivery system separated at the tip of the inner catheter during procedures.
Cook has received 13 complaints of the delivery tip separation worldwide, including one U.S. report, with an occurrence rate of 0.043 percent. Two adverse events were reported in Japan in...